Biological Indicators

What are Biological Indicators?

Biological indicators (BIs), as defined by ANSI/AAMI and ISO, are test systems containing viable microorganisms providing a defined resistance to a specific sterilization process. A biological indicator provides information on whether necessary conditions were met to kill a specified number of microorganisms for a given sterilization process, providing a level of confidence in the process. Endospores, or bacterial spores, are the microorganisms primarily used in BIs. They are considered some of the toughest ones to kill. Additionally, bacterial spores are chosen for a specific sterilization process based on their known resistance to that process. For example, Geobacillus stearothermophilus spores demonstrate a high resistance towards steam and vaporized hydrogen peroxide and are therefore used in BIs that monitor these sterilization processes.

Benefits of Biological Indicators in Sterilization Process

Biological indicators, depending on the specific type, can be used for various sterilization processes using steam, hydrogen peroxide gas, ethylene oxide and more.

Several factors such as operator experience, load preparation and sterilizer condition can impact the sterilization cycle. BIs provide a direct measure of the lethality of the process and so the use of BIs to routinely monitor sterilizers provides assurance in the efficacy of the sterilization process.

BIs are typically used within process challenge devices (PCD) that are designed to represent the most challenging products routinely processed. A passing result for the BI within this defined challenge demonstrates that the sterilizer is effective in killing a large number of highly resistant bacterial spores, providing users with a level of assurance in their sterilization process.

Biological Indicator Testing Procedure

A biological indicator is made up of a carrier material, on which bacterial spores with a defined resistance to the sterilization process have been applied. The carrier material is enclosed within a glassine envelope or a vial. The BI is exposed to the sterilization process and then incubated under defined growth conditions to determine whether any spores survived the process. If no spores survive, none grow and the test is a pass. If growth is detected, the test is a fail.1

As per AAMI ST79:20132, PCDs containing biological indicators should be used for routine monitoring of steam sterilization cycles at least weekly, preferably daily, and in every load containing implants. Sterile Processing Department staff should follow the manufacturer's directions for the appropriate placement of the PCD containing the BI in the sterilizer.

Besides routine sterilizer monitoring, PCDs containing BIs are also used for other applications as listed below:

  • Routine load release of nonimplant loads
  • Routine load release of implant loads. AAMI ST79:2013 recommends using a PCD containing a BI and a Class 5 integrating indicator. Sterile Processing staff can also use a PCD containing a BI and a Class 6 challenge pack. In either case, loads containing implants should be quarantined until the BI results are known, unless it is an emergency situation.
  • Qualification of sterilizers after installation, relocation, major repair, malfunction, and sterilization process failures
  • Periodic product quality assurance testing

Regardless of the type of process or application, Sterile Processing staff should always follow the manufacturer's instructions for use (IFUs) when using a biological indicator to monitor sterilization.

Biological Indicators for Steam Sterilizer (Autoclave)

Biological indicators (within a PCD) are often used for routine monitoring, qualification and load monitoring of a steam sterilizer. Biological indicators are designed to demonstrate whether the conditions during a steam (autoclave) cycle were adequate to achieve a defined level of microbial inactivation. For steam sterilization, BIs are typically used within PCDs and AAMI ST79:2013 recommends weekly, preferably daily, routine monitoring of sterilizers and in every load containing implants.

The PCD containing the BI and/or a chemical indicator should be placed in the most challenging location in the chamber. In a steam sterilizer, this is typically on the bottom shelf near the drain. Upon completion of the cycle, the PCD is taken out and the BI is incubated. After incubation the BI result is read. Incubators that are also readers will provide a final pass or fail result within a specified amount of time.

Types of Biological Indicators for Sterilization Processes

There are various types of biological indicators depending on the need and sterilization process:

  • Self-contained biological indicators (SCBI) – These indicators combine the carrier material with spore and growth medium into a single vial, eliminating the need for aseptic transfer.1 Following sterilization, the vial is activated allowing the spores to mix with the growth medium and incubated to allow spore growth. Some SCBI designs use an enzyme based detection system that can detect spore growth sooner that most on the market, with shorter incubation times.

Irrespective of the design, biological indicators are used during a steam sterilization process, hydrogen peroxide sterilization process or ethylene oxide process, to ensure the efficacy and lethality of the process.

Where to Buy Biological Indicators?

Healthcare facilities should purchase BIs from a reliable, accessible source for easy reordering. Biological indicators can be purchased from companies that offer sterilization and infection prevention products. Factors that may be considered when deciding which biological indicators to purchase include incubation time, brand name, cost, range and variety of products, validated applications and ease of use. Refer to ISO 14161 for more information on supplier selection.

1 https://university.steris.com/course/indicators-disinfection-and-sterilization/

2 ANSI/AAMI ST79:2013 Comprehensive guide to steam sterilization and sterility assurance in health care facilities. Association for the Advancement of Medical Instrumentation; 2013

Categories

Sterile Processing

Surgical Equipment

Email Signup

Sign up to receive updates on the latest news, events, educational opportunities and promotions from STERIS.

Sign Up

 

 
STERIS Corporation

STERIS is a leading provider of infection prevention
and other procedural products and services.

5960 Heisley Road, Mentor, OH 44060 U.S.A.

800.548.4873 or 440.354.2600

QUICK LINKS

Careers

Investors

News

 
Code of Ethics

steris.com

© Copyright 2018, STERIS plc. All rights reserved.
Registered in England and Wales No. 09257343.
Registered office: Rutherford House, Stephensons Way,
Derby DE21 6LY, United Kingdom