STERIS Corporation - Helping to provide a healthier today and a safer tomorrow.
Comparing Options for Rapid Turn Sterilization | STERIS

Comparing Options for Rapid Turn / Immediate Use Sterilization

Supporting a Compliant Approach When You Need To Sterilize a Device Quickly

Surgical procedures do not always go as planned. An instrument arrives in the operating room (OR) with a tear in the sterile packaging, or a unique instrument is dropped, and the procedure cannot continue without it. Mistakes happen, and undeniably there are instances where a device needs to be reprocessed faster than normal, in some cases, immediately.

Flash Sterilization Methods

The lack of sufficient inventory or late delivery of instrumentation should not be considered an emergency.

To accomplish this, you will often hear two terms used: Rapid Turn and immediate-use steam sterilization (IUSS). Rapid Turn is often used in the Sterile Processing Department to identify instrument sets needed the same day, moving them ahead in normal queue. IUSS, formerly known as “flash sterilization,” is defined in the standards as the shortest possible time from an item being removed from a steam sterilizer to the aseptic transfer onto the sterile field1.

The IUSS process is intended to sterilize a single instrument for emergencies only. There was a time when facilities went beyond the original intent making IUSS more common as opposed to an exception. Whatever the reason causing the need for a device to be processed immediately, it has been established by organizations (like Joint Commission) who perform audits of healthcare facilities that IUSS should not be used except in the case of a true emergency.

What do the Standards Say?

Facilities should always follow evidence-based guidelines and national standards when performing IUSS.

CDC, AORN, AAMI

Organizations, including the CDC, AORN, and AAMI, agree that IUSS should only be used in emergencies and kept to a minimum.

Why? Consider the risks associated with IUSS, which can include:

  • The lack of timely biological indicators (BI) results monitoring performance
  • Compliance/adherence to manufacturers IFUs for cleaning
  • Cooldown period required after sterilization, with possible risk of staff or patient burns
  • Minimal sterilization cycle parameters (i.e., time, temperature)2 which do not meet device instructions for use (IFUs)

However, there are emergency situations where IUSS may be the only option, and facilities should specify what constitutes an emergency in their policy and procedures. Staff should understand evidence-based guidelines and undergo competency verification to ensure the sterilization process used is successful.

It is important to follow the device or the sterilizer equipment manufacturer’s written IFUs for reprocessing, including confirming the specific cycle’s time and temperature are approved for that device. The same critical reprocessing steps (cleaning, decontamination, and transporting sterilized items) must be followed regardless of the specific sterilization cycle employed; a safe process does not include shortcuts or work-arounds4.

The IUSS process must always be documented. Documentation includes, but is not limited to, load and cycle identification (time, temperature, and load contents), BI results, and case information (operating surgeon, patient, and the reason for IUSS).

Each instance of IUSS will likely go under review during an audit. Credentialing organizations like the Joint Commission and DNV will likely review IUSS logs during every inspection. Typically, auditors check for completeness of documentation to ensure IFUs were followed and that the cleaning, decontamination, and inspection steps were entirely performed prior to IUSS. If an auditor identifies a trend that IUSS is performed frequently, they will also likely review a facility's processes and policy, questioning if lack of inventory, and not a true emergency, resulted in the need to perform an IUSS cycle.

Options When You Need to Sterilize an Item Quickly

What options exist when a hospital or ambulatory surgery center (ASC) needs to sterilize an item quickly?

Delaying Surgical Procedures

Cancel/Delay the Case
and Reschedule

Could cause Patient Risk

Though not ideal, facilities can cancel and/or delay surgeries and reschedule for a later date.

Alternative surgical devices used during surgical procedures

Seek Alternative Devices
May conflict with the surgeon’s preference

Many surgeons have preferences for specific devices they use during a surgical procedure. Another option is using an alternative brand camera, scope, drill, etc...

Immediate Use Steam Sterilization

Immediate Use Steam Sterilization
Subject to Additional Auditing

  • Due to limited cooldown time, there is a risk of patient and/or staff burns from hot devices or instruments.
  • Items in the tray come out wet due to shortened dry time. Wet items increase the chances of rust or corrosion occurring.
  • Devices are subjected to shorter steam exposure and minimal or no dry time, which can conflict with the device manufacturer’s IFUs.
  • If an instrument or device is not used after IUSS, it must go through the sterile processing process again, including decontamination, cleaning, inspection, and sterilization before re-use or storage.
Terminal Sterilization

Terminal Sterilization
The Compliant and Reliable Option

Terminal sterilization is the best option:

  • Follows standard operating procedures and guidelines.
  • Surgical instruments should be sterilized using terminal sterilization cycles within rigid sterilization containers, wrappers, or primary packaging designed to maintain the instruments’ sterility, which allows the devices to be stored for later use4.
  • Staff can be confident that the process is compliant with standard procedures.

Rapid Turn Terminal Sterilization

So, how can you remain compliant when you need a device or instrument quickly sterilized and do not have the time for a standard steam cycle?

Steam Flash Sterilization

Steam

  • IUSS is not Terminal Sterilization
  • A typical terminal sterilization cycle is approximately 1 hr. total
  • Cooldown time after sterilization could be an additional 2 - 3 hrs. depending on the metal mass
VH2O2 Flash Sterilization

VH2O2

  • Fast Cycle Options
  • Reliable
  • No Cool Down Time
  • Immediate Release
  • No special utilities are required at location of the equipment

To achieve terminal sterilization in Steam requires a standard cycle time. A typical standard cycle time for 4-minute sterilization and 30-minute dry time is approximately 1 hour total. Some IFUs require a 45 minutes or longer dry time, which lengthens the process.

In addition to the cycle and dry time, adequate cooling time can require two (2) hours or more5, depending on the conditions and package contents. Considering those times, it is easy to understand that a delay of hours is not acceptable when a device is needed quickly.

Vaporized Hydrogen Peroxide (VH2O2) Low Temperature Sterilization. VH2O2 sterilizers offer fast cycle options and ensure terminal sterilization in as little as 16 minutes for properly prepared medical devices. With VH2O2, no cool-down time is necessary because VH2O2 sterilization operates at a low temperature. Personnel can use devices or instruments immediately following their policies and procedures for release.

What if the device IFU does not specifically mention VH2O2 Sterilization?

If the device falls within the claims in the VH2O2 sterilizer’s indications for use, the VH2O2 sterilizer may be used for reprocessing. Customers are encouraged to reference the sterilizer’s Device Compatibility Matrix for a full list of compatible devices. The decision to process these devices with a sterilizer, such as V-PRO Low Temperature Sterilization Systems, would be the type of informed decision recommended by AAMI ST81, pg. B6.

Skip the risks of IUSS and protect the safety of your patients, staff, and devices by using VH2O2 Fast Cycles. VH2O2 sterilization offers the peace of mind you need during rapid turn sterilization.

Explore V-PRO Low Temperature Sterilizers

v-pro low temperature sterilization

Article References

1 https://www.jointcommission.org/standards/standard-faqs/ambulatory/infection-prevention-and-control-ic/000002122/

2, 3, 4 Flash Sterilization – Guideline for Disinfection and Sterilization in Healthcare Facilities (2019), CDC https://www.cdc.gov/infectioncontrol/pdf/guidelines/disinfection-guidelines-H.pdf

5 ANSI/AAMI ST79:2017, Section 10.3.1 – Unloading sterilizers having a chamber volume larger than 2 cubic feet.

6 ANSI/AAMI ST81 Sterilization of medical devices—Information to be provided by the manufacturer for the processing of resterilizable medical devices

Subscribe to our Mailing List

FOLLOW US
Go to LinkedIn 12,900 FOLLOWERS
Go to Facebook 8,300 FANS
Go to YouTube 2,710 SUBSCRIBERS
Go to X 3,600 FOLLOWERS
Go to Instagram 850 FOLLOWERS

 

Live Chat
Go to Top of Page